Validity, tolerability, and feasibility of automated arterial occlusion pressure across body positions
Physiological measurement · 2026 · Vol 47(8) · 085009
Dr. Nicholas Rolnick · Co-author
Abstract
Objective. Accurate arterial occlusion pressure (AOP) measurement is essential for individualized blood flow restriction (BFR) prescription, but no automated BFR device has been validated across the range of body positions commonly used in practice. This study evaluated the validity, tolerability, and feasibility of automated algorithm-based AOP measurement compared to pulsed-wave Doppler ultrasound across supine, seated, and standing positions.
Approach. Forty-six healthy adults completed AOP measurements on the dominant lower limb using both automated and Doppler ultrasound methods at three body positions in randomized order. Validity was assessed using intraclass correlation coefficients (ICC), Bland–Altman analysis, and proportional bias testing. Tolerability was evaluated through measurement duration, pain perception (0–10), and adverse event surveillance. Feasibility was assessed by the first-attempt success rate.
Main results. The automated method demonstrated good agreement with Doppler ultrasound in supine (ICC = 0.76), seated (ICC = 0.87), and standing (ICC = 0.83) positions. However, 95% limits of agreement ranged from −38 to +48 mmHg across positions, and a significant negative proportional bias was observed in supine (p = 0.031). Automated measurements required significantly longer duration than Doppler across all positions (p < 0.001), while pain perception was comparable between methods. A single participant withdrew during standing automated measurement due to intolerable pain. The first-attempt success rate was significantly lower for automated measurement in standing compared to Doppler (87% vs 100%, p = 0.041) but comparable in supine and seated.
Significance. The automated device demonstrates good validity and acceptable tolerability for AOP assessment in healthy young adults during seated and standing positions. Clinicians should consider manual ultrasound verification for supine measurements or when prescribing high-percentage AOP protocols, and provide postural support during standing measurements to improve first-attempt success.
Authors
Jinghui Yang, Taylor Gubler, Elizabeth Kenneally, Caden Cloutier, Erin Duffy, Zach Hamilton, Grace Harty, Nicholas Rolnick, Ben Senra, Drew Martin
Dr. Nicholas Rolnick’s contribution: Co-author
Keywords
- Rehabilitation
- Doppler ultrasound
- Validity
- Blood Flow Restriction
- Automated Arterial Occlusion Pressure
Read the full paper
Cite this paper
Jinghui Yang, Taylor Gubler, Elizabeth Kenneally, Caden Cloutier, Erin Duffy, Zach Hamilton, Grace Harty, Nicholas Rolnick, Ben Senra, Drew Martin. (2026). Validity, tolerability, and feasibility of automated arterial occlusion pressure across body positions. Physiological measurement, 47(8), 085009. https://doi.org/10.1088/1361-6579/ae9034
Related research
More from this line of work
Journal article · 2026
Measurement properties and operational feasibility of an automated arterial occlusion pressure assessment system for blood flow restriction exercise
Read the record
Letter / commentary · 2026
Are we comparing blood flow restriction outcomes using equivalent equipment? The importance of empirically validated equipment and accurate reporting for comparison of outcomes related to safety and efficacy
Read the record
Journal article · 2026
Autoregulation during blood flow restricted exercise offers no additional benefit on thigh muscle hypertrophy and strength adaptations in trained participants: a randomized within-subject 8-week trial
Read the record
Apply the research
From the paper to the patient
Every protocol in The Complete BFR Certification cites the literature it came from, including this line of work. The module-by-module bibliography (Bonus 5) maps each claim back to its paper.